Sulfur Hexafluoride (SF6) Trace Gas Sensing using Modulated CO2 Laser Beam

Main Article Content

Mohammed A. Hussain
Sarah S. Ahmed

Abstract

Photonic Crystal Fiber Interferometers (PCFIs) are widely used for sensing applications. This work presents the fabrication and the characterization of a relative humidity sensor based on a polymer-coated photonic crystal fiber that operates in a Mach- Zehnder Interferometer (MZI) transmission mode. The fabrication of the sensor involved splicing a short (1 cm) length of Photonic Crystal Fiber (PCF) between two single-mode fibers (SMF). It was then coated with a layer of agarose solution. Experimental results showed that a high humidity sensitivity of 29.37 pm/%RH was achieved within a measurement range of 27–95%RH. The sensor also showed good repeatability, small size, measurement accuracy and wide humidity range. The RH sensitivity of the sensor has a significant dependence on the thickness of the coating and the sensor with the highest sensitivity showed a linear response for RH change in the range of 27-95% RH and a fast response time of 0.8 sec for an RH change from 50% to 90%.

Article Details

How to Cite
“Sulfur Hexafluoride (SF6) Trace Gas Sensing using Modulated CO2 Laser Beam” (2017) Iraqi Journal of Laser, 17(A), pp. 9–15. doi:10.31900/ijl.v17iA.28.
Section
Articles

How to Cite

“Sulfur Hexafluoride (SF6) Trace Gas Sensing using Modulated CO2 Laser Beam” (2017) Iraqi Journal of Laser, 17(A), pp. 9–15. doi:10.31900/ijl.v17iA.28.

Publication Dates